Device specific registers – In-Situ RDO PRO-X Process Optical Dissolved Oxygen Sensor Manual User Manual

Page 21

Advertising
background image

800-446-7488

21

www.in-situ.com

Device Specific Registers

Register

Size

Mode &

Access Level

(R/W)

Data Type

Description

Dissolved Oxygen Concentration

0038

2

R1

float

Measured value, C

o

0040

1

R1

ushort

Parameter ID = 20

0041

1

R1/W2

ushort

Units ID
117 = mg/L (default)
118 = ug/L

0042

1

R1

ushort

Data quality ID

0043

2

R1/W3

float

Offline sentinel value (default = 0.0)

0045

1

R1

16 bits

Available units = 0x0030 (48)

Temperature

0046

2

R1

float

Measured value

0048

1

R1

ushort

Parameter ID = 1

0049

1

R1/W2

ushort

Units ID
1= °C (default)
2= °F

0050

1

R1

ushort

Data quality ID

0051

2

R1/W3

float

Offline sentinel value (default = 0.0)

0053

1

R1

16 bits

Available units = 0x00030 (3)

Dissolved Oxygen % Saturation

0054

2

R1

float

Measured value

0056

1

R1/W2

ushort

Parameter ID = 21

0057

1

R1/W2

ushort

Units ID
177 = percent saturation (default)

0058

1

R1

ushort

Data quality ID

0059

2

R1/W3

float

Offline sentinel value (default = 0.0)

0061

1

R1

16 bits

Available units = 0x0001 (1)

Oxygen Partial Pressure

0062

2

R1

float

Measured value

0064

1

R1

ushort

Parameter ID = 2 (pressure)

0065

1

R1/W2

ushort

Units ID
26 = torr (default)

0066

1

R1

ushort

Data quality ID

0067

2

R1/W3

float

Offline sentinel value (default = 0.0)

0069

1

R1

16 bits

Available Units = 0x0200 (512)