In-Situ RDO PRO-X Process Optical Dissolved Oxygen Sensor Manual User Manual

Page 5

Advertising
background image

800-446-7488

5

www.in-situ.com

Optical RDO PRO-X Dissolved Oxygen Probe Specifications

Optical RDO PRO-X Dissolved Oxygen Probe

Sensor Type

Optical (luminescent) dissolved oxygen sensor

RDO PRO-X Probe

Range: 0 to 50 mg/L concentration

Accuracy: ±0.1 mg/L, 0 to 8 mg/L

±0.2 mg/L, 8 to 20 mg/L

±10% of reading, 20 to 50 mg/L

Resolution: 0.01 mg/L

Response time: T90 < 45 sec; T95 < 60 sec @ 25°C

Storage conditions: -5° to 60° C (23° to 140° F)

RDO-X Sensor Cap

Usage life: 2 years typical

Storage conditions: 1° to 60° C (33° to 140° F), in factory container

Temperature sensor

Range: 0° to 50° C (32° to 122° F)

Accuracy: ±0.1° C typical

Resolution: 0.01° C

Transmitter/local display

Optional, or use with Con TROLL PRO System

Communications options

Modbus/RS485, 4-20 mA, SDI-12

Max. power consumption

50 mA at 12 VDC

Measure current

6 mA typical at 24 VDC

Idle current (no measurement
or communication)

160 µA typical at 24 VDC

Maximum cable length

Up to 1219 m (4000 ft) (for Modbus and 4-20 mA) or up to 60.96 (200 ft) (for SDI-12)

Cable options

10 m fixed stripped-and-tinned; Twist-lock connector with user-selectable cable length

Internal mounting thread

1¼ - 11 ½ NPT

IP rating

IP-67 with cap off, IP-68 with cap installed

Compliance

Heavy industrial, IEC 61000-6-2:2005

Salinity compensation

Fixed or real-time capable (using controller)

Barometric pressure

Fixed or real-time capable (using controller)

Maximum pressure

150 psi from 0 to 50° C; 300 psi @ 25° C

Warranty

Probe: 3 years from date of shipment
RDO-X Sensor Cap: 2 years in typical applications

Methods

Standard Methods 4500-O; In-Situ Methods 1002-8-2009,1003-8-2009, 1004-8-2009
(EPA approved)