5.0 Controller Requirements and Connections

The RDO PRO Probe may be connected to a controller or logger for communication via:

Analog (4-20 mA) provides a configurable 4-20 mA current loop output

SDI-12

RS485 Modbus

RS232 Modbus

5.1

Wiring Overview

Refer to diagrams on the following pages. Trim and insulate unused wires. The shielded wire should
be wired to a chassis ground or earth ground.

Signal

Color

Ground/Return

Black

External Power

Red

4-20 mA

Brown

RS485 (-)

Green

RS485 (+)

Blue

SDI-12

White

The inside of the controller must be kept free of moisture and humidity. Condensed
moisture can migrate through the wiring and cause the probe to fail. Therefore,
desiccant should be installed in the controller and be replaced on a regular basis.