Calibrate the depth sensor, Zero in air – In-Situ smarTROLL Multiparameter Handheld Operators Manual User Manual
Page 44
Advertising

44
If your calibration standard references 20° C, tap the
Thermometer icon and change the reference temperature.
6. Once the calibration is stable, tap the Accept button.
7. To view the calibration report, tap View Report.
8. Rinse the sensors with DI water.
Calibrate the Depth Sensor
Zero in Air
1. Tap the Calibration icon
to access a list of sensors that are available for
calibration.
2. Tap Depth Sensor.
3. Tap Zero in Air.