Calibrate the depth sensor, Zero in air – In-Situ smarTROLL Multiparameter Handheld Operators Manual User Manual

Page 44

Advertising
background image

44

If your calibration standard references 20° C, tap the
Thermometer icon and change the reference temperature.

6. Once the calibration is stable, tap the Accept button.

7. To view the calibration report, tap View Report.

8. Rinse the sensors with DI water.

Calibrate the Depth Sensor

Zero in Air

1. Tap the Calibration icon

to access a list of sensors that are available for

calibration.

2. Tap Depth Sensor.

3. Tap Zero in Air.