Aqua troll, Accessories – In-Situ RuggedCable for Aqua TROLL—Male connector end + stripped and tinned end User Manual

Page 2

Advertising
background image

1 800 446 7488

(toll-free, US and Canada) or 970 498 1500

www.in-situ.com

Due to continuing product development this information is subject to change without notice. In-Situ and the In-Situ logo, Win-Situ, TROLL, BaroTROLL,
RuggedReader, RuggedCable and RDO are trademarks or registered trademarks of In-Situ Inc. Viton is a registered trademark of DuPont Dow Elastomers.
Copyright © 2005-2007 by In-Situ Inc. All rights reserved.

0053312 rev. 004 09/07

Aqua TROLL

®

Accessories

SHORT STRIPPED & TINNED CABLE

Aqua TROLL Connections

MODBUS MASTER

with RS485 built in

*

Optional but highly recommended

EXT PWR RED

GND/RETURN BLACK

Digital PLC

12-36 VDC

*

RS485 (–) GREEN

RS485 (+) BLUE

Signal

Color

Gnd/Return

BLACK

Ext Power

RED

4-20 mA

BROWN

RS485(–)

GREEN

RS485(+)

BLUE

SDI-12

WHITE

SDI-12 3 WIRE

EXT PWR RED

GND/RETURN BLACK

Data Logger

9.6-16 VDC

SDI-12 WHITE

RS485 (–) GREEN

RS485 (+) BLUE

MODBUS MASTER

with RS232 built in (converter required)

*

Voltage limited by converter

EXT PWR RED

GND/RETURN BLACK

Digital PLC

12 VDC

*

RS485 (–) GREEN

RS232 (RXD)

RS232 (TXD)

Converter

Gnd +12V

RS485 (+) BLUE

RS485

RS232

***

Required if port power is not available

DB9F

TD(A)

TD(B)

GND

GND

+12V

RS485 (+) BLUE

RS485 (–) GREEN

PIN

5

GND

PIN

3

TXD

PIN

2

RXD

GND/RETURN BLACK

***

EXT PWR RED

***

CONVERTER

Port-Powered RS485 converter,
such as B&B Electronics Model
485SD9TB

ANALOG (4-20 mA) 3 WIRE

4-20 mA BROWN

PLC or Data Logger

GND/RETURN BLACK

+

-

SIGNAL

12-36 VDC

EXT PWR RED

Advertising