In-Situ Quick-Cal Check Solution User Manual

Quick cal solution, Information sheet

Advertising
background image

Information Sheet

Due to continuing product development this information is subject to change without notice. In-Situ, the In-Situ logo, and TROLL are trademarks or regis-
tered trademarks of In-Situ Inc. Copyright © 2002–2006 by In-Situ Inc. All rights reserved.

1 800 446 7488

(toll-free, US and Canada) or 970 498 1500

www.in-situ.com

Multi-Parameter

Water Quality TROLL

®

Catalog No. 33250

0032825 rev. 003 3/06

QUICK CAL
SOLUTION

pH, ORP, & Conductivity values

at various temperatures

pH

mV

µS/cm

Temp.

(± 0.02)

(± 5)

(± 40)

5

o

C

7.10

255

4910

10

o

C

7.04

247

5635

15

o

C

7.03

239

6390

20

o

C

7.02

231

7180

25

o

C

7.00

224

8000

30

o

C

6.98

217

8850

35

o

C

6.97

209

9735

Specific Conductivity Coefficient:

1.95% /

degree C (0-25

o

C)

2.06% /

degree C (15-35

o

C)

2.23% /

degree C (25-50

o

C)

APPLICATION
A calibration standard of an appropriate chemical composition to perform a one-
point calibration of all four MP TROLL Basic sensors (pH, ORP, Conductivity,
Polarographic Dissolved Oxygen) at the same time with a single solution. Four
250 mL containers, each sufficient for at least 2 Quick Cals.

Shelf Life, unopened

6 months

After opening

7-21 days

EXPECTED ACCURACIES
The following accuriacies may be expected when calibrating the MP TROLL
using Quick Cal solution and following instructions in the Win-Situ Quick Cal
Wizard and/or in the MP TROLL Operator's Manual:

GUIDELINES AND PRECAUTIONS

•

Store the solution cool and dark. Shake well before use. For best results,
calibrate at the temperature expected during field use

•

Do not store sensors in Quick Cal solution

•

For best results, use fresh solution for each Quick Calibration

•

Dispose of used solution by flushing down the drain with plenty of water

•

Solution is best if used before date on bottle, but may be usable beyond that
date if stored cool, dark, and tightly capped

Quick Cal

Meets USGS

Parameter

accuracy

guidelines*

pH: at pH 7

± 0.09 pH units

✓

✓

✓

✓

✓

over range 0-14 pH

± 0.2 pH units

✓

✓

✓

✓

✓

ORP

± 25 mV

NA

DO (with stirrer)

± 0.25 mg/L

✓

✓

✓

✓

✓

Conductivity:
Low Range (3 – 50,000 µS/cm)

± 3%

✓

✓

✓

✓

✓

High Range (70 – 200,000 µS/cm) ± 3%

✓

✓

✓

✓

✓

* Calibration criteria for continuous water-quality monitors, from USGS Water-Resources
Investigations Report 00-4252, p. 13