5 calibration solution fill line, Instruction sheet – In-Situ smarTROLL Multiparameter Handheld Calibration & Storage Cup User Manual

Page 2

Advertising
background image

Instruction Sheet

3. Set the instrument in the calibration cup.

Do not fasten the instrument to the cylinder.
The chamber must be vented to the
atmosphere to produce a valid calibration.

4. Follow the calibration directions in the Operator's

Manual or in the Mobile Help to complete the
procedure.

Calibration Solution Fill Line

1. Calibration procedures that require calibration

solutions should be performed after you have
completed the 100 percent oxygen saturation
calibration.

Important: Remove the sponge before you
add solution to the calibration cup.

2. Fill the calibration cup to the top of the interior metal

sleeve.

3. Set the instrument in the calibration cup

4. Follow the calibration directions in the Operator's

Manual or in the Mobile Help to complete the
procedure.

2

Advertising