Instruction sheet – In-Situ TROLL Net Hub Instruction Sheet User Manual

Page 3

Advertising
background image

Instruction Sheet

Number

Wire Color

Function

3

Blue

RS485 (+)

4

Red

Power DC+

The center screw terminal block is the master, which
connects to the TROLL

®

Link Telemetry System or

PLC. On 8-port TROLL Net Hubs, which has two circuit
boards, one circuit board master screw terminal is wired
to the second circuit board master screw terminal.

Number

Description

1

Board 1 Master Terminal (to board 2
Master)

2

Board 2 Master Terminal (to telemetry,
PLC, or other TROLL Net Hub)

For bulkhead connector models, disconnect the Red
Power DC+ wire from the screw terminal of any port
that is not being used.

NEVER disconnect the Red Power DC+ wire
from a master terminal.

Internally (9V Battery) Powered 4-port or
8-port TROLL Net Hub

Each screw terminal of an internally powered TROLL
Net Hub is wired the same way, with the exception of
the master terminal(s).

Number

Wire Color

Function

1

Black

Ground

2

Green

RS485(-)

3

Blue

RS485(+)

The center screw terminal block is the master, which
connects to the TROLL

®

Link Telemetry System or

PLC. On 8-port TROLL Net Hubs, which has two circuit
boards, one circuit board master screw terminal is wired
to the second circuit board master screw terminal.

Number

Wire Color

Function

1

Black

To 9V Battery (-)

2

Black

Ground

3

Green

RS485(-)

4

Blue

RS485(+)

5

Red

To 9V Battery (+)

DO NOT connect the red Power DC(+) wire from an
instrument cable to a screw terminal when the
TROLL Net Hub is internally powered. Connecting this
wire causes the instrument to use the 9V battery as an

800-446-7488

3

www.in-situ.com