In-Situ Flow Cell for Low-Flow Measurements (2-in diameter sonde) User Manual

Page 2

Advertising
background image

1 800 446 7488

(toll-free, US and Canada) or 970 498 1500

www.in-situ.com

Multi-Parameter Water Quality TROLL

®

This information is subject to change without notice. In-Situ and the In-Situ logo, Win-Situ, Pocket-Situ, TROLL, RuggedCable, RuggedReader,
and RDO are trademarks or registered trademarks of In-Situ Inc. Teflon is a registered trademark of E.I. DuPont de Nemours and Company.
Copyright © 2002 – 2010 by In-Situ Inc. All rights reserved.

0044712 rev. 002 04/10

5. Attach hoses to valves and secure with clamps (may not be

needed with very low flow rates).

6. To provide stability, attach spike to base of flow cell and press

into the ground. Can be secured at a small angle from vertical
(< 30°) to allow bubbles to self-purge from the cell.

7. Remove the restrictor and insert the calibrated MP TROLL into

the cylinder. Tighten threads to attach (not too tight).

physiCAL DesCription
Dimensions

6.4 cm (2.5 in) O.D., 21.4 cm
(8.4 in) long

Volume

200 mL without sensors

Materials

Clear polished polycarbonate

Ground spike

316L stainless steel

NPT attachment

3

/

8

” NPT

Hose sizes accepted

3

/

8

”, ½”

Typical flow rate

0.1 - 0.5 liters/minute

Recommended

Flow rate: 8.0 liters/minute

operational limits

Pressure: 30 psi

ACCessories

CAtALog no.

Flat base 12" x 12" PVC, with hardware .................... 44430
Flow-Sense™ low-flow software .....................................NA

(included in Win-Situ 4 and Pocket-Situ 4, version 4.40 and later)

Hardware fittings kit ................................................... 44340

(contains 4 barbed adapters, 6 hose clamps, pipe thread tape)

guiDeLines AnD preCAutions
• Flow is controlled by the pump, not the 3-way ball valve. The

valve may be used to divert flow for sample collection.

Low -FLow CeLL

Catalog No. 0044710

• To decrease air bubbles in the system, tape the adapter fittings

before installation.

• To prevent interference with sensors, ensure that adapter fit-

tings do not protrude beyond the inside wall of the flow cell.
Replacement

3

/

8

” NPT adapter fittings are available from In-Situ

Inc. readily available at hardware stores and plumbing supply
houses.

• In-Situ's Flow-Sense low-flow software calculates volume auto-

matically.

• Clean and decontaminate the inner stainless steel sleeve after

use. Re-install, either side up.

to pump

Check valve

Ball valve

Optional
turbidity
sleeve

Spike

(or base plate)

Flow

Flow

Advertising