For more information contact In-Situ Inc.

221 East Lincoln Avenue, Fort Collins, CO 80524
1-800-446-7488 (toll-free in U.S. & Canada)
1-970-498-1500 (international & domestic)
www.in-situ.com

0032005 rev. 019 09/11

Information subject to change without notice. In-Situ, In-Situ logo, Baro Merge, BaroTROLL, HERMIT, Pocket-Situ, RDO, RuggedCable, RuggedReader, TROLL, and Win-Situ are trademarks or registered trademarks of In-Situ

Inc. © 2011. All rights reserved. Tefzel is a registered trademark of E.I. du Pont de Nemours & Co. Viton is a registered trademark of DuPont Performance Elastomers L.L.C.

Potential Interferences

pH

Sodium salts

Nitrate NO

3

-

HCO

3

-

hydrogen carbonate*

CO

3

2-

carbonate

CIO

4

-

perchlorate

I

-

iodide

CIO

3

-

chlorate

CN

-

cyanide

Br

-

bromide

NO

2

-

nitrite

HS

-

hydrogen sulfide

CI

-

chloride

H

2

PO

4

-

hypophosphate

AcO

-

acetate

F

-

fluoride

SO

4

2-

sulfate

*Significant interference above pH3

Ammonium NH

4

+

Cs

+

cesium

K

+

potassium

TI

+

thallium

Na

+

sodium

Li

+

lithium

H

+

pH

Ag

+

silver

Chloride Cl

-

Complexes with:
Bi

3+

bismuth

Cd

2+

cadmium

Mn

2+

manganese

Pb

2+

lead

Sn

2+

tin

Tl

3+

thallium

Reducing agents:
OH

-

hydroxide

NH

3

ammonia

S

2

O

3

2-

thiosulfate

Br

-

bromide

S

2-

sulfide

I

-

iodide

CN

-

cyanide

DO (Clark Cell)
CO

2

, HS

-

, O

3

DO (Optical)
NaCIO (bleach) >3%
Alcohols >5%
H

2

O

2

(peroxide) >3%

Gaseous SO

2

Gaseous Cl

2

Do not use in organic solvents, e.g.,
acetone, chloroform, or methylene chloride.

ORP

Ions that are stronger reducing agents

than hydrogen or platinum, e.g.,

chromium, vanadium, titanium

Calibration Solutions

Solution

Shelf Life After Opening

Quick Cal (pH, ORP, conductivity)

7-21 days (±10 mV, ±0.05 pH, ±50 µS)—store cool and dark, shake before use

ZoBell’s (ORP)

7-21 days (±10 mV)—store cool and dark

Low Conductivity (143 µS)

hours (±1 µS check before use)

Other Conductivity

3-6 months

pH Buffers

3-6 months

Sodium Sulfite

3-6 months

Turbidity

12 months

Low Chloride

1 month

Low Nitrate

1 month

Low Ammonium

1 month

Other ISE Solutions

2 months

pH Fill Solution

6-12 months

TROLL

®

9500 Sensor Summary