In-Situ Nose Replacement Kit for the optical RDO PRO Probe User Manual
Instruction sheet, Nose replacement kit for the rdo, Pro and rdo

Instruction Sheet
Nose Replacement Kit for the RDO
®
PRO and RDO
®
PRO-X Probe
Part Number 0080820
Kit Contents
1
Replacement Nose Cone
2
Installed O-ring
3
O-ring for probe body
4
Lubricant
5
Lens cloth
Replacement Procedure
1. Remove the nose cone and existing O-ring from the
probe body.
1
O-ring
2
Nose Cone
2. Place a small amount of lubricant on your fingertip.
Coat the entire O-ring with lubricant. Do not allow any
dust or debris to get on the O-ring.
3. Place the new O-ring on the probe body.
4. Use the lens cloth to remove any lubricant from the
probe and cap.
5. Place the nose cone on the probe body.