Nose Replacement Kit for the RDO

®

PRO and RDO

®

PRO-X Probe

Part Number 0080820

Kit Contents

1

Replacement Nose Cone

2

Installed O-ring

3

O-ring for probe body

4

Lubricant

5

Lens cloth

Replacement Procedure

1. Remove the nose cone and existing O-ring from the

probe body.

1

O-ring

2

Nose Cone

2. Place a small amount of lubricant on your fingertip.

Coat the entire O-ring with lubricant. Do not allow any
dust or debris to get on the O-ring.

3. Place the new O-ring on the probe body.

4. Use the lens cloth to remove any lubricant from the

probe and cap.

5. Place the nose cone on the probe body.