SECTION 2 – Installation

2.1

UNPACKING

Remove the 6300 or 6320D from the packaging and ensure the following items are
included:

1. Model 6300 spectrophotometer, or Model 6320D spectrophotometer

2. Required power cable either UK mains lead moulded plug (HH179(S)), EU mains lead

moulded plug (HH180(S)), or US mains lead moulded plug (FCABLEUS)

3. Pack of 100 disposable plastic visible wavelength cuvettes (060 084), model 6300 only

4. Jenway 63-zero PC software (635 100) and interface cable (013 210) model 6300 only

5. Instruction manual (630 026)

6. Jenway Foreign Manual CD (JENMANCD)

7. Optional accessories (as ordered)

2.2

INSTALLATION

Models 6300 and 6320D are supplied ready to use.

The unit should be placed on a clean flat surface which is free from drafts and vibrations. The
units are designed for operation on 115V to 230V AC input at 50 to 60Hz. The standard 2 metre
cable supplied with the units is fitted with an IEC type connector which can be plugged directly
into the POWER IN socket on the rear panel. The unit should be positioned within 1.5 metres of
an earthed mains supply.

The mains fuse is housed within the POWER IN socket. When replacing the fuse the unit must be
disconnected from the mains supply. The fuse rating is 2A ‘F’ (fast blow type). In the event of the
fuse failing after replacement it is advisable to consult with Bibby Scientific Service or your
distributor before proceeding further.

It is possible to switch the voltage between 230V and 115V. When changing the voltage please
ensure the fuse rating is correct. Before attempting to change the voltage disconnect the
instrument from the mains supply. Withdraw the fuse holder from the power input socket and
remove the fuse. Extract the grey fuse retainer and rotate so that the correct voltage is visible
through the aperture in the fuse holder. Replace the fuse retainer in its holder, fit the correct fuse
and push assembly back into the power input socket. For further information please refer to the
service manual.

Connect the power supply unit to the power inlet socket on the rear panel of the instrument and
connect to the mains socket. Turn the power on at the mains and switch the instrument on using
the power switch on the rear of the instrument. DO NOT open the sample chamber door during
initialisation.