SECTION 9 – Troubleshooting

9.1 ERROR CODES

Refer to the table below to diagnose error codes. If you require further assistance contact Jenway
Help (refer to section 9.2).

Error Code

Error Name

Issue

Err 1

Dark cal error


This error occurs when the 6300/6320D closes the shutter to

block light entering the sample chamber. If the detector

output does not fall to a level normally associated with a

dark cal then this error is indicated. The most likely cause is

that the sample chamber lid is not closed.

Err 2

Light cal error




This error indicates that there is insufficient light to calibrate

to 100%. The most likely cause is that the light at the

selected wavelength is being absorbed by a sample in the

sample chamber.

This error can also be caused if the lamp has failed.

Err 3

Concentration
calibration error

The calculated factor is outside the instrument range.

Recalibrate the instrument using a standard of lower

concentration.

Err 4

Lamp Failure

The lamp failed after initially finding 0nm peak. If the lamp

has failed, it should be replaced.

Err 5

No light detected


The instrument has failed to find a zero order peak. Possible

causes of this error are that the lamp has failed or that a

sample has been left in the unit which is absorbing the zero

order light.

Err 6

No dark level
detected

The instrument checks the region below 320nm which

should have no optical throughput. If this region cannot be

found the most likely cause is that the sample chamber lid is

not closed.

9.2

TECHNICAL SUPPORT

Jenway have a dedicated Technical Support team made up of experienced scientists who are on
hand to help with any applications advice and questions you may have about our products

and

how to use them. If you require any technical or application assistance please contact the team
at:

E-mail:

[email protected]

Phone: +44 (0)1785 810433