KANOMAX 1560 Multi-Channel Anemomaster Software Manual User Manual

Page 5

Advertising
background image

4

3. Functional Specification

ITEM CONTENTS

Model No.

1550, 1560, 1500

Module

V, VT, VTH, SP

Numbers of
ANEMOMASTER

5 units (maximum) can be connected.

Channel Nos to measure MAX. 320CH (in case of V module) CH Nos depend on module.
Probe positioning

Optional position in 3 dimension of rectangular coordinates.

Measurement mode

Channel mode

Data mode

Average value mode, Instant mode

Interval of measurement 0.1~6553.5seconds, you can set up by 0.1 second unit
Repetition time

65535 times MAX.

Port

COM1、COM2、COM3、COM4

Baud rate

300, 600, 1200, 2400, 4800, 9600, 19200 bps

Delimiter

CRLF

, CR

Length of data 8 bit
Parity None

RS232C

Stop bit

bit

Statistical calculating
function

calculate maximum, minimum, average value and standard deviation from data

of each probe

Parameter

display

Display of module set up, probe set up, RS232C parameter, measurement

parameter

Time serial
graph

Data for one probe is displayed in one window.

MAX.320 of time series graph windows can be opened simultaneously.

Select probe by table.

Vertical :V,T,H,P (by probe) , scale can be set up optionally.

Horizontal : times

Switch scroll and pages

Display of statistic data

Display

Data table

Display of all channel data value of one sampling simultaneously.

Display of module, channel, probe condition, probe nos, comment, probe position,

data ( possible to select yes or no to display unit indication ).

Possible to select indication items other than data.

Time series
graph

Print out graph in active “time series graph window”.

Print

Data table

Print out work sheet of data table.

User back up function

Version information

Tool hint

Status indication

Indication of file information and comment input