Cleaning the probe – KANOMAX 6332D Airflow Transducer User Manual

Page 21

Advertising
background image

14

Cleaning the Probe


If any impurities such as dust, soot and oil are attached to the sensor, the amount of heat that the airflow
draws from the sensor surface (heat dissipation rate) will change, thus the air velocity reading will be
affected.
The probe must be cleaned when the sensor is contaminated from use in an unclean environment.


Clean the sensor portion of the probe in an ultrasonic cleaner for approx. 10-20 sec. Do not clean the

sensor longer than required as excess cleaning will lead to sensor damage.

Use tap water (or purified water) for cleaning. The sensor portion can also be cleaned in a vessel

filled with neutral detergent diluted with water, and by stirring it in the vessel.



!) Make sure to TURN OFF the power before cleaning.
!) Dry the probe completely after cleaning. Do not use the probe before it is completely dried.

Cleaning Method

Clean in an ultrasonic cleaner for approx.
10-20 sec.

Do not point the sensor upwards until it is
completely dried. The water drops may get
inside the probe and cause malfunction.

! Caution !

Dry with cold air.