4-4

BHK-MG (OPR) 022014

4.3.3

LOW CURRENT CALIBRATION

1. LCD reads

CONNECT SHUNT <nn OHM >zz A

(nn and zz values are determined by BHK-

MG model being calibrated.; e.g., for BHK 2000-0.1MG, LCD reads

<50 OHM >10 W

).

(Refer to PAR. 4.2 for shunt requirements.) Connect Precision Shunt across power output
terminals (TB1, terminals 2 and 5) and press ENTER.

2. LCD reads

CONNECT DVM TO SHUNT

. Connect DVM (+) input to sense terminal of shunt

connected to output (+), DVM (–) to sense terminal of shunt connected to output (–). Press
ENTER when connections are complete.

3. With LCD reading

Curr ZERO adj.

, monitor DVM and use

or

keys to increase or

decrease output current as necessary for minimum positive DVM reading. Press ENTER
when complete.

4. LCD will flash

Curr F.S.adj. **WAIT**

then

Curr F.S. adj < >

. Monitor DVM and

use

or

keys for fine adjustment and 3 or 1 keys for coarse adjustment to increase or

decrease output current as necessary to get V

DVM

reading as close as possible to value

listed below.

V

DVM

= Rs x Io

MAX

(mV)

where

Io

MAX

(mA) is the maximum value of output current (low scale):

BHK 300-0.6MG:

0.6/10 = 0.06A = 60mA

BHK 500-0.4MG:

0.4/10 = 0.04A = 40mA

BHK 1000-0.2MG:

0.2/10 = 0.02A = 20mA

BHK 2000-0.1MG:

0.1/10 = 0.01A = 10mA

Rs (Ohms) is the actual value of the external shunt resistor (the same shunt resistor can be
used for both Low and High scale current calibration).

NOTE: The message

Can’t Adj. Full Scale

is displayed if either the shunt resistor is not

connected to the output terminals during current calibration or if the shunt resistance is
too high. If this occurs, it is necessary to press ENTER followed by RESET to restore
proper operation. Connect the proper shunt across the power output terminals (TB1,
terminals 2 and 5) and refer to PAR. 4.3, step 1 to restart calibration.

5. LCD will show

Curr 10x adj

then

value =____mA

. Monitor DVM and convert V

DVM

to

mA as follows, then note the value of Io

10%LFS

(milliAmps).

Io

10%LFS

(mA) = V

DVM

(mV)/Rs (ohms)

where

Io

10%LFS

(milliAmps) is the 10% full scale value of low range output current (1% of I

OMAX

)

Rs (Ohms) is the actual value of the external shunt resistor (the same shunt resistor can be
used for both Low and High scale current calibration).

6. Use the keypad to enter the value of Io

10%LFS

(milliAmps) noted in step 5 and press ENTER.

If the value is out of range, the message

Out of range

then

Measure again

will be dis-

played. Repeat steps 5 and 6 until directed to proceed to High Current calibration.

7. Press ENTER when complete and proceed to PAR.4.3, step 6.