Table 3-10. square waveform frequency vs. points, 5 viewing stored waveforms, Viewing stored waveforms -27 – KEPCO BOP 1KW-MG Operator Manual, Firmware Ver.2.38 to 2.47 User Manual

Page 91: 10 square waveform frequency vs. points -27

Advertising
background image

BOP HIPWR 080709

3-27

3.3.9.5

VIEWING STORED WAVEFORMS

After observing the precautions of PAR. 3.3.9.1, press

@

from the power-up screen to access

the Saved Waveforms screen (Figure 3-9). Use

Y

or

U

or the encoder to highlight the name of

a previously saved waveform, then press

!

to view the Waveform Settings screen (see Figure

3-10). The lower half of the LCD shows a representation of the programmed waveform. The let-
ters along the vertical axis of the graph at the lower left side indicate whether the waveform is
programmed to operate in voltage or current mode. The message at the upper right indicates
the actual mode of the power supply while the waveform is being executed. The upper half pres-
ents the waveform name, positive and negative protection settings, the repetition count, fol-
lowed by a listing of segments. The type (square, sine, etc.), frequency, p-p amplitude, and

TABLE 3-9. SINE, TRIANGLE AND RAMP WAVEFORM FREQUENCY VS. POINTS

Frequency

(See Notes 1, 2, and 3)

Total Points

Frequency

(See Notes 1, 2, and 3)

Total Points

From

To

From

To

0.01Hz

2.7Hz

3840

55.5Hz

66.5Hz

160

2.71Hz

3.6Hz

2880

66.6Hz

88.7Hz

120

3.71Hz

5.5Hz

1920

88.8Hz

118.3Hz

90

5.6Hz

8.3Hz

1280

118.4Hz

147.9Hz

72

8.4Hz

11.0Hz

960

148Hz

177.4Hz

60

11.1Hz

14.7Hz

720

177.5Hz

221.8Hz

48

14.8Hz

22.1Hz

480

221.9Hz

295.8Hz

36

22.2Hz

33.2Hz

320

295.81Hz

354.9Hz

30

33.3Hz

44.3Hz

240

355.9Hz

443.7Hz

24 (See Note 2)

44.4Hz

55.4Hz

192

443.8z

532Hz

20 (See Note 3)

NOTES: 1. As the frequency varies within the range, the time interval per point varies proportionately.

2. Sine and Triangle segments can not exceed 443Hz.
3. Ramp segments can not exceed 532Hz.

TABLE 3-10. SQUARE WAVEFORM FREQUENCY VS. POINTS

Frequency

(See Note 1)

Total Points

Frequency

(See Notes 1)

Total Points

From

To

From

To

0.02Hz

1.8Hz

3840

43.51Hz

58.0Hz

120

1.81Hz

2.7Hz

2880

58.01Hz

72.5Hz

90

2.71Hz

4.0Hz

1920

72.51Hz

87.0Hz

72

4.01

5.4Hz

1280

87.01Hz

108.7Hz

60

5.41

7.2Hz

960

108.71Hz

145.0Hz

48

7.21Hz

10.8Hz

720

145.2.9Hz

174.0Hz

36

10.81

16.3Hz

480

174.1Hz

217.5Hz

30

16.31Hz

21.7Hz

320

217.6Hz

261.0Hz

24

21.71Hz

27.1Hz

240

261.1Hz

435.0Hz

20

27.11Hz

32.6Hz

192

435Hz

652.5Hz

12

55.5Hz

66.5Hz

160

653

1000

10

NOTE: 1. As the frequency varies within the range, the time interval per point varies proportionately.

Advertising