2 calibration procedure using local mode, Caution, If the calibrate zero – KEPCO BOP 1KW-MG Operator Manual, Firmware Ver.2.01 to 2.37 User Manual

Page 136: Calibration procedure using local mode -12, R. 4.4.2

Advertising
background image

4-12

BOP-1K 022108

%

or

ENTER

- EXIT - Exit Voltage or Current Calibration

4. Pressing

!

,

@

,

#

, or

$

displays following explanations for the function keys:

CAUTION: IF THE CALIBRATE ZERO

!

FUNCTION IS AVAILABLE, PERFORM THIS

FUNCTION FIRST, AS IT AFFECTS ALL OTHER OUTPUT CALIBRATIONS.

!

- Calibrate the Zero adjustment for this operational mode. IF THIS FUNCTION IS

AVAILABLE, PERFORM THIS FUNCTION FIRST, AS IT AFFECTS ALL OTHER OUT-
PUT calibrations.

@

- Adjust the MAXIMUM (Positive) for this operational mode.

#

- Adjusts the MINIMUM (Negative) for this operational mode

$

- PROTECT - Adjusts External Protection limits. (Active for EXTERNAL screen only.)

%

or

ENTER

to exit this menu.

5. Selecting one of the above options begins calibrating the output. The screen describes the

function of the active keys:

!

- or clockwise rotation of the

ADJUST

control adjust the output by approximately 10

increments in the positive direction.

@

- or clockwise rotation of the

ADJUST

control while ppressed adjusts the output by

one increment in the positive direction.

#

- or counterclockwise rotation of the

ADJUST

control adjusts the output by 10 incre-

ments in the negative direction.

$

- or counterclockwise rotation of the

ADJUST

control adjust the output by approxi-

mately 10 increments in the negative direction.

%

- Exits the active calibration and allows you to select other calibrations that are

required.

Y

- Increments the output D to A by 100, causing the output to increase.

U

- Decrements the output D to A by 100, causing the output to decrease.

4.4.2

CALIBRATION PROCEDURE USING LOCAL MODE

NOTES:

1. For ZERO calibration, adjust output to obtain value closest to zero.

2. For FULL SCALE calibration adjust the output to obtain the closest value above

(absolute value) the nominal full scale value.

3. Set the unit for local sensing with GND Network in (see Figure 4-1) and discon-

nect load from BOP output.

4. When adjustment is needed refer to PAR. 4.4.1, step 5.