B.24 [SOURce:]CURRent[:LEVel][:IMMediate][:AMPlitude] COMMAND
CURR
Syntax:
Short Form: [SOUR:]CURR[:LEV][:IMM][:AMP] <NR2>
Long Form: [SOURce:]CURRent[:LEVel][:IMMediate][:AMPlitude] <NR2>
Where <NR2> = current setpoint
Description: Current Setting. The current setpoint affects the load only when the load is in Current mode of opera-
tion. Current setpoint is stored in non-volatile memory; value is restored upon power-up.
B.25 [SOURce:]CURRent[:LEVel][:IMMediate][:AMPlitude]? QUERY
CURR?
Syntax:
Short Form: [SOUR:]CURR[:LEV][:IMM][:AMP]?
Long Form: [SOURce:]CURRent[:LEVel][:IMMediate][:AMPlitude]?
Returns <NR2> = current setpoint
Description: Indicates Current Setpoint.
B.26 [SOURce:]CURRent:PROTection[:LEVel] COMMAND
CURR:PROT
Syntax:
Short Form: [SOUR:]CURR:PROT[:LEV] <NR2>
Long Form: [SOURce:]CURRent:PROTection[:LEVel] <NR2>
where: <NR2> = current protection setpoint
Description: Establishes current protection setpoint. Current protection setpoint is stored in non-volatile mem-
ory; value is restored upon power-up. This is a user-defined value which, if exceeded during opera-
tion, causes a FAULT condition, forcing the load to disengage.
B.27 [SOURce:]CURRent:PROTection[:LEVel]? QUERY
CURR:PROT?
Syntax:
Short Form: [SOUR:]CURR:PROT[:LEV]?
Long Form: [SOURce:]CURRent:PROTection[:LEVel]?
Returns: <NR2> = current protection setpoint
Description: Indicates current protection setpoint.
B.28 [SOURce:]CURRent:PROTection:STATe COMMAND
CURR:PROT:STAT 0
Syntax:
Short Form: [SOUR:]CURR:PROT:STAT 0
Long Form: [SOURce:]CURRent:PROTection:STATe 0
Description: Clears current protection level error.
B.29 [SOURce:]CURRent:PROTection:STATe? QUERY
CURR:PROT:STAT?
Syntax:
Short Form: [SOUR:]CURR:PROT:STAT?
Long Form: [SOURce:]CURRent:PROTection:STATe?
Returns: <bool> = 1 (current protection error occurred)
0 (current protection error has not occurred)
Description: Indicates whether a current protection error has occurred (1) or not (0).
B.30 [SOURce:]MODE COMMAND
MODE
Syntax:
Short Form: [SOUR:]MODE <txt>
Long Form: [SOURce:]MODE <txt>
where: <txt> = CURRent, POWer, VOLTage, RESistance, CONDuctance, SHORT or OFF.
Description: Sets operating mode: When the operating mode is changed, the input is set to off. Send INPut ON to
apply the new mode setting to the load.
B.31 [SOURce:]MODE? QUERY
MODE?
Syntax:
Short Form: [SOUR:]MODE?
Long Form: [SOURce:]MODE?
Returns: <txt> = CURRent, POWer, VOLTage, RESistance, CONDuctance, SHORT or OFF.
Description: Indicates operating mode.