Installation, 4 storage, 5 pre-installation requirements – KROHNE BATCHFLUX 5500 C Quickstart EN User Manual

Page 6: 6 general requirements

Advertising
background image

2

INSTALLATION

6

BATCHFLUX 5500 C

www.krohne.com

05/2014 - 4003386301 - QS BATCHFLUX 5500 R03 en

2.4 Storage

• Store the device in a dry and dust-free location.
• Avoid lasting direct exposure to the sun.
• Store the device in its original packaging.
• Storage temperature: -50 ...+70°C / -58...+158°F

2.5 Pre-installation requirements

Make sure that you have all necessary tools available:

• Small wrench (M5) for connection to ground
• Torque wrench for installing flowmeter in pipeline

Assecoires necessary for the correct installations are available on request at the manufacturer

Make sure that these accesoires are available before starting installation;

• O- ring / L-ring gasket
• Special pipe flanges
• Stud bolt with lockwasher, plain washer and nut

2.6 General requirements

INFORMATION!
To facilitate servicing and/or exchanging of the device, please note that:
it must be possible to shut off the flow through the pipeline ( control valve upstream in pipeline).
Drain the pipeline before removing device ( provide drain valve)

INFORMATION!
The following precautions must be taken to ensure reliable installation.

•

Make sure that there is adequate space to the sides.

•

Protect the signal converter from direct sunlight and install a sun shade if necessary.

•

Support the pipeline on both side of the flowmeter.

•

Do not expose the signal converter to intense vibration. The flowmeters are tested for a
vibration level in accordance with IEC 60068-2-64.