Grounding ring number 1 (optional for DN25...150):
• Thickness: 3 mm / 0.1" (tantalum: 0.5 mm / 0.02")
Note:
Note:
Note:
Note: For diameter DN10 and DN15, grounding rings are integrated as standard in the flow
sensor construction.
3.3 Virtual reference for IFC 300 (C, W and F version)
The virtual reference option on the IFC 300 flow converter provides complete isolation of the
measurement circuit.
Benefits of virtual reference:
• Grounding rings or grounding electrodes can be omitted.
• Safety increases by reducing the number of potential leakage points.
• The installation of the flowmeters is much easier.
Minimum requirements:
• Size: ≥ DN10
• Electrical conductivity: ≥ 200 µS/cm
• Electrode cable: max. 50 m / 164 ft, type DS
3.4 Connection diagrams
Figure 3-2: Grounding ring number 1
Figure 3-3: Virtual reference
INFORMATION!
For the connection diagrams please refer to the documentation of the applicable signal
converter.