Installation – KROHNE OPTIFLUX 4000 Quickstart EN User Manual
Page 10

2
INSTALLATION
10
OPTIFLUX 4000
www.krohne.com
04/2010 - 7309852300 - QS OPTIFLUX 4000 R03 en
2.5.11 Pump
2.5.12 Temperatures
Figure 2-14: Installation after pump
CAUTION!
Protect the device from direct sunlight.
Temperature range
Process [°C]
Ambient [°C]
Process [°F]
Ambient [°F]
min.
max.
min.
max.
min.
max.
min.
max.
PTFE
Separate flow sensor
-40
180
-40
65
-40
356
-40
149
Compact with IFC 300
-40
140
-40
65
-40
284
-40
149
Compact with IFC 100
-40
140
-40
65
-40
284
-40
149
PFA
Separate flow sensor
-40
180
-40
65
-40
356
-40
149
Compact with IFC 300
-40
140
-40
65
-40
284
-40
149
Compact with IFC 100
-40
140
-40
65
-40
284
-40
149
ETFE
Separate flow sensor
-40
120
-40
65
-40
248
-40
149
Compact with IFC 300
-40
120
-40
65
-40
248
-40
149
Compact with IFC 100
-40
120
-40
65
-40
248
-40
149
Hard rubber
Separate flow sensor
1
-5
80
-40
65
23
176
-40
149
Compact with IFC 300
1
-5
80
-40
65
23
176
-40
149
Compact with IFC 100
1
-5
80
-40
65
23
176
-40
149
PU
Separate flow sensor
-5
65
-40
65
23
149
-40
149
Compact with IFC 300
-5
65
-40
65
23
149
-40
149
Compact with IFC 100
-5
65
-40
65
23
149
-40
149
1 Hard rubber liner is available for Ex-versions only