4.4 Earthing and equipotential bonding
The signal converter shall be connected to the equipotential bonding system of the hazardous
area via the external grounding connection on the signal converter housing. The measuring unit
and the signal converter are electrically connected via an equipotential bonding conductor.
Any existing cable shields should be connected to earth according to applicable installation
regulations (EN 60079-14). A terminal connection in the terminal compartment permits a short
way earthing of the cable shields.