Temperature limits, 2 ufc 300 f(/i)-ex – KROHNE OPTISONIC 6300 EEx EN User Manual
Page 12

2
TEMPERATURE LIMITS
12
OPTISONIC 6300
www.krohne.com
03/2010 - 7309972300 - EX OPTISONIC 6300 R03 en
2.2 UFC 300 F(/i)-Ex
The UFC 300 F(/i)-Ex ultrasonic signal converter in separate version is not influenced by the
temperature of the process medium, because it is installed on a distance of the pipe-line and
thus not physically connected to the pipe-line. The UFC 300 F(/i)-Ex signal converter has a
temperature classification of T6 (85°C). The permissible ambient temperature is dependent on
the material that the electronics housing is made of, namely:
• die-casted aluminum: -40...+60ºC
• die-casted stainless steel: -40...+55ºC
2.3 OPTISONIC 6000 xxxxxx-Ex
The OPTISONIC 6000 xxxxxx-Ex clamp-on ultrasonic flow sensor has the following maximum
process temperatures at the maximum ambient temperature T
a
of 70°C.
Temperature class
Temperature class
Temperature class
Temperature class
Maximum process temperature [
Maximum process temperature [
Maximum process temperature [
Maximum process temperature [ºC] at T
C] at T
C] at T
C] at T
a
aa
a
= 70ºC
T6
80
T5
95
T4
120
1
T3
195
2
T2
200
2
1 XT version: 130°C
2 XT versions only