KROHNE DA Flap-type EN User Manual
Page 13
Definition of Min1 and Min2 /
Max1 and Max2
Min1
Min2
Max1
Max2
8.1.4
Technical data of limit switches
SC3,5-N0-Y
SJ3,5-SN
SJ3,5-S1N
SB3,5-E2
2-wire
2-wire
2-wire
3-wire
NAMUR
NAMUR
NAMUR
Switching element function
NC contact
NC contact
NO contact
NO contact PNP
Nominal voltage U0
8 V
8V
8V
10 to 30 V
Power consumption:
Pointer vane not detected
≥ 3 mA
≥ 3 mA
≤ 1 mA
≤ 0.3 V
Pointer vane detected
≤ 1 mA
≤ 1 mA
≥ 3 mA
Ub - 3 V
Continuous current
-
-
-
max. 100 mA
No-load current I0
-
-
-
≤ 15 mA
An isolating switching amplifier, e.g. Pepperl + Fuchs Series KF .. -SR2 ..., is required in order to
operate
the
SC3,5-N0-Y
limit switch (refer to the chapter on the spare part list).
SJ3,5-SN
and SJ3,5-S1N limit switches safety-oriented are connected to a safety-oriented isolating
switching amplifier, e.g. Pepperl + Fuchs K… –SH- ... (large S on the front)
current
consumption
with
the
shown
pointer
position
SC3,5-N0
≤1 mA
SJ3,5-S1N
≤1 mA
current
consumption
with
the
shown
pointer
position
SJ3,5-S1N
≥3 mA
SC3,5-N0
≥3 mA