Installation, 2 how to assemble the segmented coaxial probe – KROHNE OPTIFLEX 1100C Quickstart EN User Manual

Page 7

Advertising
background image

INSTALLATION

2

7

OPTIFLEX 1100 C

www.krohne.com

01/2012 - 4001226201 - QS OPTIFLEX 1100 R01 en

2.2.2 How to assemble the segmented coaxial probe

• Put the signal cable into the first segment of tube. Make sure that the signal cable goes

through the center hole in the spacer.

• Attach the first segment of tube to the thread at the top of the signal cable. Tighten with a

17 mm open-end wrench.

• Attach the lock screw to the top of the tube. Tighten with a 3 mm Allen wrench.
• Do these steps again for each segment of tube, but do not attach the lock screw to the last

segment of tube.

Attach the tube segments

Figure 2-5: How to assemble the segmented coaxial probe: part 1

1 Top segment of tube
2 Bottom segment of tube
3 1 HC M6×5 lock screw
4 Bonded plastic spacer

CAUTION!

•

After you tighten the lock screw on each segment of tube, it will not be possible to
disassemble the coaxial probe.

•

If it is necessary to adjust the probe length, do not attach the lock screw to the last segment
of tube!