2.8 Electrical data
BW25/.././M9./../
BW25/.././M9./../
BW25/.././M9./../
BW25/.././M9./../…/..
/..
/..
/..
The electronic signal output ESK and the limit switch may only be connected to intrinsically safe
circuits.
Maximum values shall be maintained for the isolated circuits and the following
reactances shall be taken into consideration:
BW25/.././M10
BW25/.././M10
BW25/.././M10
BW25/.././M10
Circuit
U
i
I
i
P
i
C
i
L
i
ESK2A
30 VDC
100mA
1W
≈20 nF
≈0 μH
ESK3-PA
24 VDC
FISCO- Field Device
SC3,5-N0-Y..
16 VDC
25mA
64mW
150nF
150 μH
I7S23,5-N...
SJ3.5-SN
16 VDC
25mA
64mW
30nF
100 μH
SJ3.5-S1N
When connecting the I/O interfaces for BW25/.././M10 liquid level indicators, the following data
shall be taken into account:
Only for connection to circuits with "protected extra-low voltage (PELV)" .
For nominal values I/O function and nominal values of the non-certified slave see BW25
handbook.