6 set up, 1 general – KROHNE OPTISWITCH 3000C 2wire EN User Manual
Page 16

6 Set up
6.1 General
The numbers in brackets refer to the following illustrations.
On the electronics module you will find the following indicating
and adjustment elements:
l
DIL switch for mode adjustment - min./max. (1)
l
Signal lamp (4)
Note:
As a rule, always set the mode before starting the setup of
OPTISWITCH 3000 C . If the instrument is used in conjunction
with a signal conditioning instrument, always set the mode
switch (1) on OPTISWITCH 3000 C to max. mode.
The mode is selected on the signal conditioning instrument
with the mode switch.
The switching output will change if you set one of the two mode
switches afterwards. This could possibly trigger other con-
nected instruments or devices.
Function/Configuration
16
OPTISWITCH 3000 C - with two-wire output
Set up
30061
-EN
-060830