Installation – KROHNE OPTISENS PH 8100 EN User Manual
Page 19

INSTALLATION
3
19
OPTISENS PH 8100
www.krohne.com
03/2012 - 4001925901 MA OPTISENS PH 8100 R01 en
3.7.3 Mounting to SENSOFIT IMM 1000 immersion holder
INFORMATION!
The immersion assembly is an optional accessory and not part of the standard scope of delivery.
CAUTION!
•
Never touch or scratch the pH sensitive glass tip of the electrode.
•
Make sure that the glass tip is clean and dust-free. If necessary, clean the tip as described on
page 39
Figure 3-10: Overview of the immersion holder
1 Immersion holder
2 Protective cage
3 Spigot nut
4 Clamps
5 Electrode
6 Cap with cable gland
7 Sensor cable
8 Sensor connector
CAUTION!
Moisture inside the sensor connector must be avoided! Moisture will shortcut the mV signal
between the electrodes and deliver erratic readings!
If moisture has entered the connector dry it with air (e.g. hair blower).
.book Page 19 Wednesday, July 4, 2012 11:07 AM