Operation – KROHNE OPTISENS PH 9100 EN User Manual

Page 28

Advertising
background image

4

OPERATION

28

OPTISENS PH 9100

www.krohne.com

09/2012 - 4002314801 - MA OPTISENS PH 9100 R02 en

After starting-up the signal converter, the measuring screen appears. This is the standard
screen which is displayed automatically in the normal operating mode. If you are in this mode
and you want to adjust the temperature compensation, you have to perform the following steps:

Step 1: activating the temperature compensation for measurement

Press > for more than 2.5 seconds, then release the button. You are on the main menu level. In the
upper line of the display "A" appears, beneath the main menu quick setup is highlighted.

Press  or  until the main menu setup is highlighted.

MAIN MENU

A quick setup

B test

> C setup

D service

Press > to enter the chosen menu.

You are on the first submenu level. In the upper line of the display "setup" and "c1" appears,

beneath the submenu process input A is highlighted.

Press  or  to select process input A or process input B. Choose process input A or B
where ever pH is configurated.

Press > to enter the chosen menu.

You are on the second submenu level. The submenu parameter is highlighted.

Press > to enter the chosen menu.

You are on the third submenu level. In the upper line of the display "parameters"

appears.

Press  or to select pH.

Press ^ to confirm the entered value.

You are on the second submenu level. The submenu parameter is highlighted.

Press  or to select temperature.

Press > to enter the chosen menu.

You are on the third submenu level. In the upper line of the display "temperature"

and "C1.15.1" or "C2.15.1" appears, beneath the option probe is highlighted.

Press  or  to select temp.comp.

Press > to enter the chosen menu.

You are on the fourth submenu level. In the upper line of the display

"temp.comp" appears.

Now you can setup the temperature compensation.

Press  or  to choose between automatic or off.

Press ^ to confirm the entered value.

You are on the third submenu level. If you have chosen the option automatic, you

can select the type of temperature compensation now.

Press  or  to choose probe.

Press > to enter the chosen menu.

You are on the fourth submenu level. Press  or  to choose between

Pt100, Pt1000, manual or process input A.

Press ^ to confirm the entered value.

Press ^ several time until you reach the measuring mode again. Choose
yes to safe and confirm your selection.

INFORMATION!
When ordering the dual channel version, the interface "Pos A" and "Pos B" are populated. A
separate temperature sensor has to be connected to "Pos A" and configurated on process input
A. To connect a sensor with integrated Pt100/Pt1000, the integrated temperature sensor has to
be connected on the same terminal block as the sensor.

Advertising