OPERATION
20
SMARTSENS ORP 8150
www.krohne.com
11/2013 - 4002546702 - MA SMARTSENS ORP 8150 R02 en
4.3 Status messages and diagnostic information
Measurements out of specification
Maintenance
Message
Message
Message
Message
Description
Description
Description
Description
Action
Action
Action
Action
S01
ORP value > ORP max
ORP value out of range
Consider the measuring range limits, otherwise
select a suitable sensor for the process conditions of
the application.
S02
ORP value < ORP min
ORP value out of range
S03
Temp. value > Temp max
Temperature value out of range
S04
Temp. value < Temp min
Temperature value out of range
Message
Message
Message
Message
Description
Description
Description
Description
Action
Action
Action
Action
M01
Offset < -99 mV
Recalibrate the sensor. For more information refer
to
Calibration with PACTwareTM on page 17
M02
Offset> +99 mV
M03
Maintenance interval expired