Operation – KROHNE SMARTSENS PH 1590 EN User Manual

Page 20

Advertising
background image

4

OPERATION

20

SMARTSENS PH 1590

www.krohne.com

11/2013 - 4002545902 - MA SMARTSENS PH 1590 R02 en

4.1.2 Calibration with HART

®

Handheld 475 FIELD COMMUNICATOR

Manual hold

• Start the function calibration

calibration

calibration

calibration in menu mode Quick Setup

Quick Setup

Quick Setup

Quick Setup or Setup

Setup

Setup

Setup.

• Activate the function manual hold

manual hold

manual hold

manual hold to avoid an alarm.

• Select Yes

Yes

Yes

Yes.

• Press Enter

Enter

Enter

Enter to proceed.

Temperature compensation

• Set the temperature compensation

temperature compensation

temperature compensation

temperature compensation for calibration.

• Select between automatic

automatic

automatic

automatic or manual

manual

manual

manual.

i

If you have selected manual temperature compensation

manual temperature compensation

manual temperature compensation

manual temperature compensation, please enter the temperature of

the pH buffer solution. Default setting: 25°C / 77°F.

• Press Enter

Enter

Enter

Enter to proceed.

Start calibration procedure

• Dip the sensor into the first pH buffer solution 1 and wait till the value is stable.
• Press Ok

Ok

Ok

Ok to proceed.

i

The measurement of pH buffer solution 1

pH buffer solution 1

pH buffer solution 1

pH buffer solution 1 starts and the values pH and temperature appear.

After approx. 10 seconds the following values appear:
pH value
temperature value
Status buffer detection → Finished
Value indentified buffer

• Press Ok

Ok

Ok

Ok to confirm the identified pH buffer solution. If neccessary change the value of the pH

buffer solution manually.

• Press Enter

Enter

Enter

Enter to proceed.

i

If the sensor does not recognise the pH buffer solution the messages
Identified buffer

Identified buffer

Identified buffer

Identified buffer → 0 pH

0 pH

0 pH

0 pH and Buffer not recognised

Buffer not recognised

Buffer not recognised

Buffer not recognised appear.

• Press Ok

Ok

Ok

Ok to proceed.

i

Enter the value of the pH buffer solution manually.

• Press Enter

Enter

Enter

Enter to proceed with the calibration procedure.

• Clean the sensor with demineralised water 2.

Figure 4-2: Calibration procedure