3 instrument repair – KROHNE SU 501 EX EN User Manual
Page 23

?
Malfunction during function test
l
After pushing the test key, the switching conditions do
not occur in the correct sequence or correct duration, e.
g. no full signal is outputted.
à
Measure the line resistance
à
If the cable is highly resistive, bring it to a normal
resistance, e.g. check terminals and cable connections
on corrosion
7.3 Instrument repair
If a repair is necessary, please proceed as follows:
You can download a return form from our Internet homepage
http://www.krohne-mar.com/fileadmin/media-lounge/PDF-
Download/Specimen_e.pdf.
By doing this you help us carry out the repair quickly and
without having to call back for needed information.
l
Print and fill out one form per instrument
l
Clean the instrument and pack it damage-proof
l
Attach the completed form and possibly also a safety data
sheet to the instrument.
SU 501 Ex - Signal conditioning instrument
23
Maintenance and fault rectification