Malfunction during function test

l

After pushing the test key, the switching conditions do

not occur in the correct sequence or correct duration, e.

g. no full signal is outputted.

à

Measure the line resistance

à

If the cable is highly resistive, bring it to a normal

resistance, e.g. check terminals and cable connections

on corrosion

7.3 Instrument repair

If a repair is necessary, please proceed as follows:
You can download a return form from our Internet homepage

http://www.krohne-mar.com/fileadmin/media-lounge/PDF-

Download/Specimen_e.pdf.
By doing this you help us carry out the repair quickly and

without having to call back for needed information.

l

Print and fill out one form per instrument

l

Clean the instrument and pack it damage-proof

l

Attach the completed form and possibly also a safety data

sheet to the instrument.