Torques and pressures
Torques and pressures
Torques and pressures
Torques and pressures
Nominal size
Pressure rating
Bolts
Max. torque [Nm]
25
PN 40
4 x M 12
73,5
40
PN 40
4 x M 16
178
50
PN 40
4 x M 16
178
80
PN 40
4 x M 16
178
100
PN 16
4 x M 16
178
Nominal size [inch]
Flange class [lb]
Bolts
Max. torque [ftlb]
1
150
4 x 1/2"
54,2
1 1/2
150
4 x 1/2"
131,3
2
150
4 x 5/8"
131,3
3
150
4 x 5/8"
131,3
4
150
4 x 5/8"
131,3
Use Stainless steel A2 / 6.9 class bolts.
Tighten the bolts in fixed order: 1+2 (in), 3+4 (out), 5+6 (in), 7+8