KROHNE VARIFLUX IFS 6000 EN User Manual
Page 7

7
5
Installation in the pipeline
L
max
L
min
•
Installation material, see page 3.
•
Pipe and operating pressures,
see Section 6 “Limits”.
•
Flange spacing
= dimension “a” + 2 x gasket thickness of D2
(dimension “a”: see Sect. 11 “Dimensions and weights”;
arrangement of gaskets: see Sect. 7 “Grounding”).
•
High-temperature pipelines
Where process temperatures exceed 100 °C/212 °F, provide for facilities to
compensate for longitudinal expansion on heat-up of the pipeline.
For short pipelines use resilient gaskets and
for long pipelines install flexible pipe elements (e.g. elbows).
•
Position of flanges
• Torques
Install flowmeter in line with the pipe
For all flanged instruments to DIN, ANSI
axis. Pipe flange faces must be
and JIS
parallel to each other,
Tighten down stud bolts and nuts uniformly
max. permissible deviation:
in the sequence (1-8) shown in the drawing.
L
max
- L
min
≤≤
0.5 mm
Max. torque:
≤
0.02”
32 Nm ~ 3.2 kpm ~ 23.1 ft x lbf
For all other connections to DIN and ISO
Tighten down to metal stop.