MagCheck

5

MagCheck

Technical Data

The responsibility as to the suitability, intended use and corrosion-resistance of the materials used in
their construction rests solely with the purchaser.

Function and applications
Automatic verification of …

●

KROHNE electromagnetic flowmeters (EMFs)
with IMoCom signal converters

●

IFC 010 K+F

●

IFC 020 K+F+E*

●

IFC 090 K+F (but not “i”)

●

IFC 110 F

●

IFC 110 PF (TIDALFLUX, partially filled,
without testing level measurement)

●

IFC 210 E*

●

with all connected primary heads of the IFS series and M 900

*

Please note: special adapter (connection) leads are required for
all systems and signal converters marked with “*” currently not
included with supply, special fabrication.

Manual testing of …

●

KROHNE signal converters, including older series*,
but excluding testing of primary heads, such as

●

T 900 F+E

●

SC 80 A+AS / SC 100 A+AS

●

IFC 080 K+F / IFC 200 / IFC 200 E / AQUAFLUX 070

●

all above signal converters with IMoCom interface

Not suitable for …

●

all systems of hazardous duty (Ex) design *

●

flowmeters K 300 / K 310 / CAPAFLUX / BATCHFLUX /
ALTOFLUX 2W and all systems of hazardous-duty (Ex) design

●

all systems with signal converters SC 150 F / F 200 (DELAFLUX) /
TIV 60 / TIV 500 (a.c. field) / IFC 090 K or F of “i” design

Scope of test
with automatic verification

IMoCom signal converter

●

Field current and field frequency

●

Primary signal processing (input amplifier,
analog-digital conversion at 25%, 50%, 75%, 100% of
set measuring range)

●

Active curent output at factory calibration points

●

Pulse output at 50% of set full scale range

●

The MagCheck PC program identifies and signals changes to the
set values between verifications that affect measuring accuracy

Primary head:

●

Electrode resistances

●

Insulation resistance of the field coils

●

Resistance of field coils

Manual testing

KROHNE signal converters only

●

Field current and field frequency
at selectable flow values (0-100% of set full-scale range,
adjustable in steps of 0.1%):

●

Primary signal processing (input amplifier, analog/digital converter)

●

Current output

●

Pulse output
additionally for IMoCom signal converters:

●

Hardware info

●

Status info of ADC, display, I/O