KROHNE ALTOFLUX IFM 6080 K EEx EN User Manual
Page 10

10
9 Technical
data
See also the standard Installation and operating instructions.
Rated pressure
See typeplate
Ambient temperature:
-20..50/60
°
C, see Certificate of Conformity or following table.
Medium temperature:
See Certificate of Conformity or following table.
Max. ambient temperature
[
°
C]
Temperature class
Max. Medium temperature
[
°
C]
50
T6
T5
T4
T3
70
85
120
180
60
T6
T5
T4
T3
70
85
100
100
Note:
The maximum medium temperatures listed in the table above are safety-technical limits! Due to
function-technical reasons lower limits may prevail.
10 Maintenance
The IFM 6080 K / i- EEx compact flowmeters are maintenance free with regard to flowmetering
properties.
Within the scope of the periodic inspections required for electrical apparatus in hazardous areas, it
is recommended to check the flameproof enclosures for external damage and signs of corrosion.
Opening of the housing of the primary head (coil housing) is not allowed, as this will nullify the
used type of protection for the explosion safety. There are also no serviceable parts inside the
primary head.