KROHNE IFC 090F Converter EEx-ATEX EN User Manual
Page 13

ALTOFLUX IFC 090 F-EEx and ALTOFLUX IFC 090 F / i-EEx
13
Terminal compartment MODIS version IFC 090i-EEx
Sticker with handling information
for insulating cover plate
Sticker indicating the intrinsically
safe signal in-/output circuits
Insulating cover plate
The PE (or FE) conductor must be connected to the press-fitted M5 clamp terminal marked inside
the terminal compartment. This conductor must be guided through the rectangular opening in the
metal dividing plate that separates the non-intrinsically safe power supply terminals from the
intrinsically safe signal in-/output terminals.
2.7 Connection
diagrams
MODIS
Sect. 2.4 shows the block diagrams of the EEx electromagnetic compact flowmeter. The power
supply (terminals 1L , 0L ) is connected via cable B. The PE terminal must be connected to the
protective earth conductor of the mains supply.
The IFC 090i-EEx electronics unit is provided with intrinsically safe signal in-/output circuits due to
the installed pair of MODIS modules in accordance with the table below.