KROHNE OPTISENS PAC 050 EN User Manual

Page 11

Advertising
background image

Manual OPTISENS PAC 050

11

OPTISENS PAC 050

2.2 Connection diagram wall-mounting enclosure


Connection Terminals

Notes
1 = reference electrode = screen

pH electrode

1 - 2

2 = measuring electrode = core

1 = measuring electrode = core

ORP electrode

1 - 2

2 = reference electrode = screen

Measurement = brown & white
1 = brown, 2 = white
Supplied voltage, yellow & green

Impedance converter

1 - 4

3 = yellow = -6V, 4 = green = +6V

Pt 100

6 - 7

Display contrast

Display

Potentiometer to adjust the display contrast

Analog output

11 + 12

11 = +, 12 = -

max. burden 500 Ohm

Relay 1

14 + 15

Relay 2

16 + 17

Relay 3

18 + 19

Alarm relay

Power supply

20+21+22

check information given on instrument label

RS 485 (option)

23 + 24

23 = 1, 24 = +

Digital input

26 + 27

26 = +, 27 = -, external controller stop or lack-of water indication

Advertising