Operation – KROHNE OPTISENS PH 8590 User Manual

Page 33

Advertising
background image

OPERATION

4

33

OPTISENS PH 8590

www.krohne.com

03/2012 - 4001931401 MA OPTISENS PH 8590 R01 en

• Press ^

^

^

^ to confirm.

i

The slope of the electrode is displayed (should be near 59 mV in an intact electrode).

• Press ^

^

^

^ to confirm.

i

The message store cal. value is displayed on the screen. The converter asks if the new
calibration values should be stored.

• Choose yes

yes

yes

yes to store the calibration values. Choose no

no

no

no to discard the results.

• Press ^

^

^

^ to confirm.

i

Prior returning to the measuring display, you are asked if the configuration should be
stored.

• Choose yes

yes

yes

yes using 







or 







to store the new calibration values.

i

You have completed the pH calibration.

• If you want to return to the measuring mode, press ^

^

^

^ several times until you reach this mode.

Step 5: re-installing the electrode

• After the calibration procedure, rinse the electrode with tap water.
• Reinstall the electrode into its assembly, refer to

Installing the sensor on page 17.

Step 6: switching back to measurement

• Deactivate the function "manual hold" again.

INFORMATION!
If an error occurs during the calibration procedure, the display shows an error message.
Possible causes for an error are:

Slope too flat.

Wrong buffer solution.

Wrong electrical connection.

Moisture in the sensor connector.

.book Page 33 Monday, March 19, 2012 2:01 PM

Advertising