Technical data – KROHNE OPTIFLUX 4000 EN User Manual
Page 28

6
TECHNICAL DATA
28
OPTIFLUX 4000
www.krohne.com
01/2011 - 4000818502 - HB OPTIFLUX 4000 R02 en
Chemical properties
Chemical properties
Chemical properties
Chemical properties
Physical condition
Electrically conductive liquids
Electrical conductivity
Water: ≥ 20 μS/cm
Non water: ≥ 1 μS/cm
Permissible gas content (volume) ≤ 5%
Permissible solid content
(volume)
≤ 70%
Installation condtitions
Installation
Take care that flow sensor is always fully filled.
For detailed information see chapter "Installation".
Flow direction
Forward and reverse.
Arrow on flow sensor indicates positive flow direction.
Inlet run
≥ 5 DN
Outlet run
≥ 2 DN
Dimensions and weights
For detailed information see chapter "Dimensions and weights".
Materials
Sensor housing
DN2.5...15: Stainless steel 1.4408
DN20: GTW-S 30
DN25...3000: Sheet steel, polyurethane coated
Other materials on request.
Measuring tube
Austenitic stainless steel
Flange
Standard: Carbon steel, polyurethane coated
Other materials on request.
Liner
Standard
Standard
Standard
Standard
DN2.5...15: PFA
DN20: PTFE
DN25...150: PFA
DN200...3000: ETFE
Option
Option
Option
Option
DN200...600: PTFE
DN200...1800: PU
DN200...3000: Hard rubber (Ex only)
Other materials on request.
Connection box
(only remote versions)
Standard: polyurethane coated die-cast aluminium
Option: Stainless steel
Measuring electrodes
Standard: Hastelloy
®
C
Option: Platinum, stainless steel, titanium, tantalum, low noise
Other materials on request.