Technical data – KROHNE OPTIFLUX 7300 EN User Manual
Page 31
Advertising
TECHNICAL DATA
6
31
OPTIFLUX 7300
www.krohne.com
01/2011 - 4001163301 - MA OPTIFLUX 7300 R01 en
Passive
U
ext
≤ 32 VDC
f
max
in operating menu set to f
max
≤ 100 Hz:
I ≤ 100 mA
R
L, max
= 47 kΩ
R
L, min
= (U
ext
- U
0
) / I
max
open:
I ≤ 0.05 mA at U
ext
= 32 VDC
closed:
U
0, max
= 0.2 V at I ≤ 10 mA
U
0, max
= 2 V at I ≤ 100 mA
f
max
in operating menu set to
100 Hz < f
max
≤10 kHz:
I ≤ 20 mA
R
L, max
= 47 kΩ
R
L, min
= (U
ext
- U
0
) / I
max
open:
I ≤ 0.05 mA at U
ext
= 32 VDC
closed:
U
0, max
= 1.5 V at I ≤ 1 mA
U
0, max
= 2.5 V at I ≤ 10 mA
U
0, max
= 5.0 V at I ≤ 20 mA
NAMUR
-
Passive to
EN 60947-5-6
open:
I
nom
= 0.6 mA
closed:
I
nom
= 3.8 mA
Passive to
EN 60947-5-6
open:
I
nom
= 0.43 mA
closed:
I
nom
= 4.5 mA
U
i
= 30 V
I
i
= 100 mA
P
i
= 1 W
C
i
=10 nF
L
i
~ 0 mH
Advertising