Technical data – KROHNE H250 M40 EN User Manual
Page 66

8
TECHNICAL DATA
66
H250 M40
www.krohne.com
11/2012 - 4000640702 MA H250 M40 R02 en
ESK4-FF
For more details see the supplementary instructions "H250 M40 Foundation Fieldbus"
ESK4-PA
For more details see the supplementary instructions "H250 M40 Profibus PA"
ESK4 process variable
Values [%] from full-scale
range
Signal output [mA]
Over range
+102.5 (±1%)
20.24...20.56
Device error identification
> 106.25
>21.00
Max. current consumption
131.25
25
Multi-drop operation
4.5
Physical layer
IEC 61158-2 and FISCO Modell
Communication standard
H1 FOUNDATION Fieldbus protokol
ITK version
5.2
Power supply
Bus supply
Nominal current
16 mA
error current
23 mA
Starting current after 10 ms
< Nominal current
Physical layer
IEC 61158-2 and FISCO Modell
Communication standard
Profibus PA Profil 3.02
PNO ID
4531 HEX
Power supply
Bus supply
Nominal current
16 mA
error current
23 mA
Starting current after 10 ms
< Nominal current
MA_H250_M40_R02_en_640702_PRT.book Page 66 Wednesday, November 28, 2012 12:23 PM