7
TECHNICAL DATA
24
DK32
www.krohne.com
04/2011 - 4000898801 MA DK32 Air purge set R02 en
DK32 - Indicator with limit switches
Approvals
Cable fitting DK32/Kx/S
M16 x 1,5
Cable diameter DK32/Kx/L
7 ... 8mm
Clamp connection DK32/Kx/S
1.5mm
2
Limit switch
SC2-N0
I7S2002-N
SJ2-SN
1
SJ2-S1N
1
Type
2-wire NAMUR
2-wire NAMUR
2-wire NAMUR
Switch element function
Normally closed
Normally closed
Normally open
Nominal voltage U
0
8VDC
8VDC
8VDC
Pointer shaft not read
≥3mA
≥3mA
≤1mA
Pointer shaft read
≤1mA
≤1mA
≥3mA
DK32 with reed contact
Switching type
bistable
Switching reproducibility
<5% of full scale value
Breaking capacity
12VA
2
Max. supply voltage
30VDC
2
Max. current
0,5A
2
1 safety oriented
2 reduced values for Ex version
Standard
Indicator
Designation
ATEX
DK32 mechanical
II2GD IIC
II3GD IIC
DK32 electrical
II2G Ex ia IIC T6
IEC Ex
DK32 electrical
Ex ia IIC T6
FM
DK32
IS/I/1/ABCD;T6
NI/I/2/ABCD;T6
S/II, III/2/FG;T6
IS/I, II, III/1/A-G
NI/II/2/ABCD
Nepsi
DK32
Ex nL IIC T1-T6
Ex nA II T1-T6