KROHNE BM 702 EN User Manual

Page 22

Advertising
background image

22

Installation and operating instructions BM 702 (06/04)

Operating temperature at flange

LP flange system: -20°C (-4°F) to +130°C (302°F)

(see also chapter 8)

V96 flange system:

Basic version: -20°C (-4°F) to +130°C (266°F)

Special version: min. -60°C (-76°F)

High temperature version, FFKM:

max. +250°C (482°F)

Kalrez 2035: max. +210°C (410°F)

FPM (Viton) or FEP-coated:

max. +200°C (392°F)

PTFE-Wave-Stick: -20°C (-4°F) to +150°C (302°F),

pressure dependent

PP-Wave-Stick: -20°C (-4°F)to +100°C (212°F)

Product temperature

Unrestricted, provided ambient temperature and flange

temperature are within the specified limits

Ambient temperature

Signal converter (T

amb

): -20°C (-4°F) to +55°C (131°F)

Microwaves

Measuring principle

FMCW Radar

Frequency range

X-Band 8.5 - 9.9 GHz

Antenna radiation angle

Type 3: ± 8°

Type 4: ± 6°

Wave-Stick: ± 9°

Ex-i current output HART® (passive)

Current

4 - 20 mA; without or with error message 3.6 mA

or 22 mA or 3.8-20.5 mA according to NAMUR NE43

Firmware version 7.12 and higher

Accuracy and linearity

0.15 %; TC=100 ppm/K

clamp voltage

> 17V (I = 4 mA); > 13V (I = 20 mA)

Digital communication

HART®

Ambient conditions

Environment class

Locations exposed direct to open-air climate,

D1 Severity in conformity with EN 60654-1

Protection category (converter)

IP66 / IP67 (equivalent to NEMA 4 and 4X)

Electrical connection
Cable entries:

1 x M 20×1.5
(delivered with 1 cable gland M 20 mm or
QUICKON

®

2-pole cutter clamp)

Terminals:

Cable cross-section 0.5-1.5 mm² (AWG 20-16)

U-clamp terminals (for PA and FE) cable cross-section max. 4 mm² (AWG 12)








Advertising