5.3.3 Device settings (DTM)
When you change the configuration of a WHG-approved device with the DTM and then store to
the device, you must confirm the change 2 times. Do the procedure that follows:
Store to the device (procedure for WHG-approved devices)
• Change the device settings.
• Click on the “Store to device” button.
i
The DTM opens a "Save" window and tells the user to make sure it is necessary to store the
changes.
• If it is necessary to store the changes, click on the “Yes” button.
i
The DTM opens a “WHG WARNING” window.
• If the device settings are correct, click on the “Store to device” button.
i
The DTM opens a second “WHG WARNING” window.
• If the device settings are correct, click on the “Store to device” button.
i
The device stores the new data to the device. Some minutes are necessary to complete this
step. End of the procedure.