KROHNE ALTOFLUX 2W IFM 4042K EN User Manual
Page 27

Part B IFC 040 Signal converter
Section 4.4
02/2001
ALTOFLUX 2W
27
Display texts
Description and settings
FCt. 2.1
TEST Q
→
Test measuring range Q
Edit
not Sure
↑
YeS Sure
↵
Operator inquiry as to whether test to be carried out?
-110.0
-100.0
-50.0
-10.0
0.0
10.0
50.0
100.0
↑
110.0
Perc.
↵
If “YeS Sure”, fixed values for the outputs can be set relative
to the full-scale range.
No setting for the outputs. When the function is terminated,
the outputs operate as before.
FCt. 2.2
HARDWARE INFO
→
Hardware information and error status
inFo.no
modul ADC
↵
x.xxxxx.
xxxx
↵
StAtUS
modul ADC
↵
xxxxxx
xxxx
↵
inFo.no
modul IO
↵
x.xxxxx.
xxxx
↵
StAtUS
modul IO
↵
xxxxxx
xxxx
↵
inFo.no
modul diSPlaY
↵
x.xxxxx.
xxxx
↵
StAtUS
modul diSPlaY
↵
xxxxxx
xxxx
↵
inFo.no
modul HART
↵
x.xxxxx.
xxxx
↵
StAtUS
modul HART
↵
xxxxxx
xxxx
↵
In the event of an error, scan and note down all information
(software number = Info Number, and status).
Important if factory needs to be consulted.
Settings are not possible here.