1.3.2 Stripping (preparation) of signal cables
Please note: The numbers in the drawings designate the stranded drain wires of signalling cables
A and B, see sectional drawings in Sect. 1.3.1.
Flow sensor
Converter
Length flow
sensor
Length Converter
mm
inch
mm
inch
a 90
3.60
a 50
2.00
b
8
0.30
b
8
0.40
c
25
1.00
d
8
0.40
d
8
0.30
e
20
0.80
e
70
2.80
Signal cable A (type DS),
double shielding
for flow sensor
Signal cable A (type DS),
double shielding
for IFC 110 F Converter
Signal cable B (type BTS),
with triple shielding (bootstrap)
for flow sensor
Signal cable B (type BTS),
with triple shielding (bootstrap)
for IFC 110 F Converter
Customer-supplied materials
W
Insulation tubing (PVC), Ø 2.0-2.5 mm (Ø 1”)
X
Heat-shrinkable tubing or cable sleeve
Y
Wire end sleeve to DIN 41 228: E 1.5-8
Z
Wire end sleeve to DIN 41 228: E 0.5-8
Signal cable A
bending radius
≥ 50 mm (≥ 2”)
Signal cable A
bending radius
≥ 50 mm (≥ 2”)
Signal cable B
bending radius
≥ 50 mm (≥ 2”)
Signal cable B
bending radius
≥ 50 mm (≥ 2”)