6-5-2DAR - Dielectric Absorption Ratio
DAR measurement is almost same to PI measurement in a sense that they test
the time course of insulation. The only difference is that DAR measurement can
get result faster than the other.

DAR =

Insulation resistance

(1 min after a start of test)

Insulation resistance

(15 or 30 sec min after a start of test)

*1

DAR

1.4 or more

1.25 - 1.0

1.0 or less

Criteria

Best

Good

Bad

Note1: DAR time is selectable: 15 or 30 sec.
How to select:

Keep the MODE button(

) pressed down and rotate the Range switch to

power on KEW3127. (DAR mark starts blinking.)

Press the UP/DOWN button (

) to switch 15 sec and 30 sec displayed at

the lower left on the LCD. Select the desirable one.

Press the ENTER button (

) and confirm the selection.

Selected DAR time is saved and kept after the instrument is powered off. To

check the time currently selected, follow the step ① described as above.

6-5-3 How to measure DAR/ PI
DAR and PI are measured automatically at continuous measurement of IR

PI/DAR

mode. Set the Range switch to any range and measure the test object continuously.

- 1 min after a start of continuous measurement:
LCD shows DAR value.
- 10 min after a start of continuous measurement:
LCD shows PI value.