3 rf output level accuracy – Anritsu AWGN MU368060A User Manual

Page 70

Advertising
background image

Section 8 Performance Test

8-4

8.3 RF Output Level Accuracy

(1) Test criteria

ALC On, output: –3 dBm, level difference from results during CW and
during modulation at noise bandwidth of 7.68 MHz: ±3 dB (excluding
when noise internal adder function is On.)

(2) Measuring instruments for test

• Power meter.........ML4803A
• Power sensor ........MA4601A

(3) Setup

Anritsu

M キ 3681A

ML4803A

Power Meter

MG3681A+

MU368060A

Sensor Input

RF Output

Anritsu

POWER METER ML4803A

ML4803A

Power Sensor

(4) Test procedures

[1] Set the communication system for the MU368060A to "NOISE1".
[2] Press the [Preset] button and then set the following parameters:

Baseband : On
Noise : 7.68 MHz

[3] Set the RF Output on the MU368060A to Off.
[4] Perform zero point adjustment and sensor sensitivity calibration

of the power meter.

[5] Set the output level of the MU368060A to –3 dBm.
[6] Set the frequency to be measured on the MU368060A.
[7] Set the sensor calibration factor for the power meter.
[8] Set the RF Output on the MU368060A to On and then measure

the output level.

[9] Set the following prameters.

Digital Modulation : On

[10] Repeat procedures from [3] to [8].
[11] Find the level difference from the results [7] during CW and [9]

during modulation.